Question:

Can the Bilibed be used in an Incubator?

Answer:

No, the bilibed is not for use in Incubators. The temperature control of the incubator may not be capable of compensating for the impact of the Bilibed. Increased humidity inside the incubator being applied for the treatment of the baby, may have a negative impact on the electric/electronic system of the Bilibed or even create an electrical shock hazard. If oxygen or other gases are used for the treatment of the baby inside the incubator, this may create a fire/explosion hazard.